Micro-CT-BarPattern
The Micro-CT Bar Pattern Chip Phantom is a perfect tool to assess in-plane and axial spatial resolution of any Micro-CT system in a direct visible manner.
Order-number: QRM-MicroCT-Barpattern |
download product data sheet: |
![]() |
![]() |
![]() |
![]() |
click image to enlarge 
The QRM MicroCT Bar Pattern Phantom was
designed to provide a tool for direct visible
measurement of spatial resolution in
Micro–CT.
The chip offers a good alternative for indirect
methods to evaluate spatial resolution in high
res X-ray imaging modalities. The phantom
comprises two chips, one orientated in-plane
and one in perpendicular (axial) orientation.
The chips are placed in a full resin cylinder
with highmechanical accuracy.
The 5 x 5 mm˛ chip contains bar- and point-
pattern with diameters from 5 to 150 µm
line/point thickness.
The depth of the structures varies between 80
and 120 µm.
The different structures on the chip are
arranged in such a way over the chip, that
spatial resolution can be evaluated in the
center as well as in the periphery of the
image/chip in a singlemeasurement.
The cylinder, including two chips, is available
in any favored diameter down to 7mm.
The phantom is available with bar pattern chips placed in resin or in air!
Specifications |
|
| Base material: | resin or cylinder with air (0.3 mm wall) |
| Diameter: | 8 mm and upon request |
| Diameter of adapter: | 32 mm and upon request |
| Length: | 40 mm |
| Weight: | 30 - 70 g |
| Material of chip: | silicon |
| Contrast | silicon / resin silicon / air |